Technical Manual: TM50567
Optica 30/40/50 series
Revision: 4.0
Date of release: 2018-12-06
54/56
www.vareximaging.com
Light field indicator-contrast
(Edge contrast setting 35 x 35 cm
at 1m SID ;
EN 60601-2-54: §203.8.102.5)
> 4 : 1
Light field indicator-Accuracy
(EN 60601-2-54: §203.8.102.6)
< 2 % SID
Linear light projection
(To align collimator with image receptor)
Laser Class 1 (EN-IEC 60825-1) (optional feature)
Laser Class 2 (EN-IEC 60825-1) (optional feature)
SID precision of measurement with
retractable tape
(EN60601-1-3: §8.5.2)
< 2 % SID
Leakage radiation
(At 1 m, X-ray beam = 150 kVp, 4 mA;
EN60601-1-3: §12.4)
≤ 0.5 mGy/h
Collimator rotation (beam axis)
Optional with rotating flange (+/- 90º)
Remote control interface
Yes (via CAN bus)
Maximum load for accessory rails
70 N
Weight
9.0 kg
Size
230 x 320 x 210 mm (w x d x h)
Operation environment
Ambient temperature: 10 – 40 °C
Relative humidity:
30 – 95 %, non-condensing
Atm. Pressure:
700 – 1060 hPa
Storage
Ambient temperature: -20 – +70 °C
Relative humidity:
30 – 95 %, non-condensing
Atm. Pressure:
700 – 1060 hPa
Product label
The product label is located on the rear-side of the collimator:
Figure 9: rear view (product label)
WARNING:
If the product label is missing or not readable contact your distributor or Varex Imaging
Nederland B.V.
The figure below shows the information provided on the product label.