Generator Switch off check
Without a patient or phantom, a radiation exposure with 80kV, 100mA, 21 s is released. The Automatic
Exposure Control must terminate the exposure in less than 100 ms.
Necessary recurrent testing
WARNING:
Before granting the Automatic Exposure Control for use on patients, check the
functionality of all AEC fields with a phantom.