background image

ALARM  ICON

WARNING

INDICATES

CAUSE

ACTION                                   

GENERAL FAULT

and

– – IN FLOW

(FLASHING)

Malfunction of 

sensor or control 

system 

Internal  

component failure

Check gas supply. If not 

corrected, disconnect patient. 

Unplug AC power, press and 

hold Run/Standby button for 

3 seconds to clear the alarm, 

send for service.

GENERAL FAULT

and

– – IN 

0

2

(FLASHING)

O₂ sensor fault

Depleted or 

defective O₂ sensor

Unplug AC power, press and 

hold Run/Standby button for 

3 seconds to clear the alarm. 

Replace O

2

 sensor. Restart unit.

GENERAL FAULT

and

– – IN TEMP

(FLASHING)

Temperature 

out of range.  

Overheating or 

temperature sensor 

malfunction.

Cannot be corrected by user: 

disconnect patient. Unplug AC 

power, press and hold Run/

Standby button for 3 seconds 

to clear the alarm, send for 

service.

 

WATER OUT

No water in dispos-

able 

water path. Gas flow 

continues without 

heating or water 

circulation.

Sterile water empty, 

or obstructed inlet 

tube.

Replace water bag or straighten 

inlet tube. Restart unit. If alarm 

persists, disconnect patient 

from therapy.

DISPOSABLE

WATER PATH 

(FLASHING)

Disposable water 

path faulty or not 

detected. Unit will 

not run.

Disposable water 

path defective, not 

properly seated or 

not installed.

If disposable water path is  

present, place unit into 

Standby,  remove and replace 

disposable patient circuit to 

reset detector. Restart unit.

BATTERY

CHARGING

(STEADY)

The internal battery backup is not fully charged. The unit would not run on 

battery for the full rated time in the event of a power failure.  No action is 

necessary.

BATTERY

(FLASHING)

The unit  is running 

in BATTERY mode.

Gas flow and 

blending continues 

without heat or 

water circulation.

AC power is 

disconnected

Reconnect AC power.

BLOCKED TUBE

(FLASHING)

High back pressure

Obstructed or 

kinked cannula/

delivery tube, incor-

rect cannula for 

flow rate or DPC 

improperly seated

Clear obstruction, check 

cannula type, re-install DPC

GENERAL FAULT ALARMS: Failures in the control or measurement systems will cause a General Fault alarm 

indicated by this icon accompanied by the Temp display showing numbers between 50 & 84 (error codes) and 

dashes in O₂ and Flow displays. When an error code is displayed, gas delivery is stopped. The user needs to monitor 

the treatment and respond to general fault alarms. General Fault alarms cannot be silenced with the mute button. 

To reset, first disconnect the unit from AC power and then press the Run/Standby Button. With the exception of O₂ 

sensor replacement, the unit must be repaired by an approved service facility.

CARTRIDGE

CANNULA TYPE

OPERATIONAL FLOW RATES

High Flow

Adult, Pediatric/Adult Small, Pediatric Small*

5-40 liters per minute (L/min)

Low Flow

Premature, Neonatal, Infant, Intermediate Infant, 

Solo, Pediatric Small*

1-8 liters per minute (L/min)

ALARM  ICON

WARNING

INDICATES

CAUSE

ACTION                                   

CARTRIDGE 

FAULT

Cartridge and/or DPC not 

detected. Unit will not run.

RUN mode: faulty 

sensor or cartridge 

not detected.

Disconnect patient. Remove 

disposable patient circuit.

Check cartridge installation. 

Check sensor windows are 

clean. 

Gas bubbles in water 

circulation. Unit

continues to operate.

Excessive gas  

diffusion through 

cartridge fibers.

Disconnect patient. Place 

unit into Standby. Replace 

disposable patient circuit 

including water path, 

cartridge & delivery tube.

Cartridge and/or DPC

not detected.

STANDBY mode:  

missing cartridge.

Remove disposable

patient circuit. Check

cartridge installation. 

CARTRIDGE 

TYPE

Indicates type of cartridge installed (low or high flow).  Not an alarm.

GAS SUPPLY

(FLASHING)

GAS SUPPLY

(CONTINUOUS 

AND FLOW 

RATE NUMERIC 

DISPLAY

FLASHES)

Gas supply pressure 

outside 4-85 psi

(28-586 kPa) range.

Unit will not operate.

Gas supply is 

disconnected or 

exhausted.

Check gas supply and correct

as necessary.

Selected flow can not

be provided from current 

gas supply.

Inlet gas pressure 

too low for 

selected flow rate.

Increase gas pressure or 

decrease flow setting.

TEMPERATURE 

NUMERIC

DISPLAY 

FLASHES

Temperature 2° > set point

User enters set 

point much lower 

than previous 

temperature.

Silence alarm and wait for 

temperature to drop.

Temperature 2° < set point

Very low water 

temperature after 

bag replacement.

Silence alarm and wait for 

temperature to rise. 

INDICATIONS, WARNINGS AND CAUTIONS

Primary Indications:
The Precision Flow® Heliox is intended to add warm moisture to breathing gases from an external source for 
administration to a neonate/infant, pediatric and adult patients in the 
hospital, subacute institutions, and home settings. It adds heat and moisture to a blended medical heliox (79% 
helium, 21% oxygen)/ oxygen mixture and assures the integrity of the precise heliox/oxygen mixture via an 
integral oxygen analyzer. The flow rates may be from 1 to 40 liters per minute via nasal cannula.

Contraindications: General:  
• Not appropriate for patients who are not spontaneously breathing, are unable to protect their airway or have anatomic or 
injury induced blockage of the nasal pathway to the nasopharyngeal space
• Not for treating OSA and snoring
• The Precision Flow® Heliox is not for field transport
• The Precision Flow® Heliox is MRI unsafe. Do not use it in an MR environment.

Additional patient monitoring including pulse oximetry is necessary if the Precision Flow® Heliox is used to give 
supplementary oxygen.

LO HI

 

 

 

 

 

Precision Flow® Heliox   Packaging contains:

Precision Flow® Heliox Unit

Instructions For Use on USB

Quick Reference Guide

Power Cord

Delivery Tube clip

O2 Sensor cell

US ONLY - Heliox and Oxygen Hoses

Содержание Heliox Precision Flow

Страница 1: ...tridge fault Gas supply fault Run Stop Status LED Run Standby button Q U I C K R E F E R E N C E G U I D E High Flow Circuit Installed This guide provides you with basic instructions on how to set up and operate the Precision Flow Heliox Before operating the Vapotherm Precision Flow Heliox please review the Instructions For Use which can be found at our website www vapotherm com and on the USB dri...

Страница 2: ...ADY The internal battery backup is not fully charged The unit would not run on battery for the full rated time in the event of a power failure No action is necessary BATTERY FLASHING The unit is running in BATTERY mode Gas flow and blending continues without heat or water circulation AC power is disconnected Reconnect AC power BLOCKED TUBE FLASHING High back pressure Obstructed or kinked cannula d...

Страница 3: ...previous temperature Silence alarm and wait for temperature to drop Temperature 2 set point Very low water temperature after bag replacement Silence alarm and wait for temperature to rise INDICATIONS WARNINGS AND CAUTIONS Primary Indications The Precision Flow Heliox is intended to add warm moisture to breathing gases from an external source for administration to a neonate infant pediatric and adu...

Страница 4: ... when the set variables are reached Place the cannula on the patient Once the unit has reached at least 33 C connect to delivery tube The unit should not be placed in Standby mode for extended periods of time For pauses in therapy keep unit in RUN mode remove cannula from the patient and set the parameters to the lowest available setting To reinitiate therapy before cannula is placed on patient cl...

Отзывы: