CAUTION
Isopropyl alcohol is flammable and may easily ignite
Caution! Hot surface!
• Follow the safety instructions of the isopropyl alcohol manufacturer!
X
Place all parts of the Cooling Unit
(except the Cap Lock) and the Filling
Chamber Screen in isopropyl alco-
hol for a maximum of 1/2 hour.
X
If necessary, wipe the Cap Lock with
isopropyl alcohol.
X
Rinse all parts under running water
for at least one minute.
X
Let all parts dry completely.
6 C L E A N I N G/H YG I E N E
6.2.3 Check the Cooling Unit Parts and the Filling Chamber Screen
X
Check all parts for damaged surfaces, cracks, softening, hardening, em-
brittlement, contamination or discoloration.
X
Discard damaged parts.
6.2.4 Assembling the Cooling Unit
CAUTION
Plant particles in the Cooling Unit
Clogging of the Cooling Unit
Inhalation of plant particles
• Insert the Screen correctly into the Cooling Unit Cap.
X
Assemble the parts in reverse order.
X
Insert the small Base Seal Ring completely into the Cooling Unit Cap.
Ö
The small Base Seal Ring is no longer visible.
X
When assembling, make sure that Seal Rings are properly seated.
X
Insert Screen in the groove provided for this purpose.