V&P Scientific, Inc.
8
Technical Note 455
Modification of Watlow Calibration Offsets (if Required)
1. The Watlow controller settings and internal calibration offsets are set at the time of
manufacture and should give accurate set temperature/internal reaction temperature
correlation (i.e., desired set temperature = 50degC, then internal reaction temperature is also
= 50degC).
2. Should you wish to access calibration offsets menu to adjust the settings, follow these steps:
a.
Touch the menu button on the front of the Watlow interface.
Figure 11. Watlow interface, Main Menu button pointed out
b.
At “
Main Menu
” screen, select “Operations”
.
c.
At “
Operations Page
” screen, select
“Inputs”
.