Front Panel
2-4
RapidVac™ Smoke Evacuator Service Manual
Filter Life Indicator
The filter life indicator provides a visual indication of the status of the life of the
filter in use. Each illuminated LED represents five hours of remaining time on the
25-hour filter. The RapidVac Smoke Evacuator tracks the time each filter has
been used allowing for an accurate display of remaining use for new or previously
used filters.
Reading the Filter Life Indicator
Install an unused RapidVac filter into the system. When the smoke evacuator is
on, all five of the filter life indicator LEDs illuminate indicating the new filter has
100% filter life. The indicator will regress through subsequent LEDs until the
final LED is illuminated indicating less than 5 hours of filter life remaining. At
one hour remaining, the LED changes to amber. When the time remaining reaches
30 minutes, the amber LED flashes slowly.
R Button
When the smoke evacuator is connected to a generator using the Generator
Interlink Cable, pressing the R button synchronizes the smoke
evacuator with the use of an electrosurgical pencil. When a pencil is activated, the
smoke evacuator activates to a preselected air flow level.
Footswitch Jack
A RapidVac Footswitch (SEA3745) is available for use with the RapidVac
Smoke Evacuator.
A footswitch may be added to any system by simply plugging it into the
footswitch jack on the front panel. The unit may be turned on or off by depressing
the footswitch pedal once for each operation. For footswitching operation, the
smoke evacuator should be on and the suction level to be activated from the
footswitch should be selected on the control panel.
Service Required Indicator
The motor powering the RapidVac Smoke Evacuator has two motor brushes that
require regular maintenance every 500 hours. When the smoke evacuator has
been in use for 500 hours, the Service Required Indicator illuminates. Replace the
motor brushes as directed in
“Motor Brush Replacement” on page 4-7. After
brush replacement, reset the system by simultaneously depressing the up- and
down-arrow air flow controls until the Service Required Indicator turns off.
Содержание RapidVac 1009803
Страница 1: ...Service Manual RapidVac Smoke Evacuator...
Страница 6: ...vi RapidVac Smoke Evacuator Service Manual...
Страница 36: ...4 10 RapidVac Smoke Evacuator Service Manual...
Страница 40: ...5 4 RapidVac Smoke Evacuator Service Manual...
Страница 44: ...6 4 RapidVac Smoke Evacuator Service Manual...