Parts Shipped with the Smoke Evacuator
1-2
RapidVac™ Smoke Evacuator Service Manual
Parts Shipped with the Smoke Evacuator
When unpacking the smoke evacuator, verify that the following parts are
included:
•
Power cord
•
User’s guide and service manual
•
Quick reference card
Upon initial receipt, inspect the smoke evacuator for dents, cracks, or damage that
may have occurred during shipment. If damage is present, notify the freight
carrier for assessment.
Additional items available from Valleylab:
•
Replacement RapidVac Filters
•
RapidVac Footswitch
•
RapidVac Remote Switch Activator
•
Generator Interlink Cable
•
Laparoscopic Smoke Evacuator Tubing Kit
•
Hoses, tubing, adapters, wands and other accessories
About the RapidVac Smoke Evacuator
The RapidVac Smoke Evacuator captures particulates and adsorbs gases from
surgical smoke. The smoke evacuator is specifically designed to improve
visibility and reduce potential health hazards associated with surgical smoke. It
can be used in both open and laparoscopic procedures.
The RapidVac Smoke Evacuator is designed with a high suction, high flow rate
centrifugal action pump. The ultra-quiet motor is used to draw the smoke from the
surgical site through the vacuum tubing and into the system where it is processed
by a series of filters. A single disposable filter is used to simplify the installation
and removal during filter changes. The filter is completely enclosed to protect the
healthcare personnel from potential contamination during filter changes.
One SEA3700 Filter contains four filtration stages to capture the smoke plume:
•
Stage 1
- A prefilter that traps and removes gross particulate and
incidental fluid
•
Stage 2
- An ULPA grade (Ultra Low Penetration Air) patented filter
captures particulates and micro-organisms down to 0.1 µm at 99.9995%
efficiency.
•
Stage 3
- The highest grade virgin activated carbon, designed specifically
for Valleylab for the removal and adsorption of odors and toxic gases
produced by burning tissue. These harmful gases may constitute a health
hazard to healthcare professionals who are subjected to prolonged
exposure. The activated carbon used in the smoke evacuation system
preferentially removes toxic organic gases (rather than water vapor) and
provides optimal odor removal.
•
Stage 4 -
An expanded foam used to trap activated carbon fines from
migrating out of the filter.
Содержание RapidVac 1009803
Страница 1: ...Service Manual RapidVac Smoke Evacuator...
Страница 6: ...vi RapidVac Smoke Evacuator Service Manual...
Страница 36: ...4 10 RapidVac Smoke Evacuator Service Manual...
Страница 40: ...5 4 RapidVac Smoke Evacuator Service Manual...
Страница 44: ...6 4 RapidVac Smoke Evacuator Service Manual...