During Surgery
4-10
Force 2 User’s Guide
Activating Accessories
Handswitching pencils
– activate by pressing either the handswitch cut or
coag button.
Monopolar handswitching forceps (for coagulation only)
– activate by closing
the forceps tines.
Bipolar handswitching forceps (for coagulation only)
– activate by closing the
forceps tines.
Monopolar footswitching accessories (e.g., endoscopes, forceps, or chuckhandles)
– activate by pressing the appropriate pedal on the monopolar footswitch.
Bipolar footswitching forceps
– activate by pressing the bipolar footswitch
pedal or activate by pressing either monopolar footswitch pedal with the
front panel footswitch selector in the bipolar mode.
Warning
When not using active accessories, place them in a holster or in a clean, dry,
nonconductive, and highly visible area not in contact with the patient. Inadvertent
contact with the patient may result in burns.
Fire Hazard:
Do not place active accessories near or in contact with flammable
materials (such as gauze or surgical drapes). Electrosurgical accessories that are
activated or hot from use can cause a fire. Use a holster to hold electrosurgical
accessories safely away from patients, the surgical team, and flammable
materials.
Caution
Do not activate the generator until the forceps have made contact with the patient.
Product damage may occur.