Troubleshooting 

Please use this table below for solving problems. If this table does not 

offer a solution contact Valley Microscope: 

[email protected]

 

Trouble

Cause

Solution

Switch on but no 
light.

Plug is unreliable

Check cables

Bulb is broken

Replace bulb

Fuse is broken

Replace fuse

Bulb is flickering.

Bulb is broken

Replace bulb

Brightness of view 
field is poor or 
uneven.

Bulb brightness is low.

Rotate potentiometer to 
increase brightness.

Objective isn’t in correct optical 
path.

Rotate the objective in 
correct position.

The size of aperture 
diaphragm is too small.

Adjust the size of 
aperture diaphragm.

Lens (objective,eyepiece, 
condenser, illuminator 
collector) has dust.

Clean lens

Position of condenser is too 
low .

Rotate condenser up.

Image  isn’t  clear 
(contrast or 
resolution is poor).

 

Cover glass thickness is off-
standard. 

Use required thickness 
cover glass (0.17mm).

Cover glass of specimen isn’t 
in up direction.

Place specimen 
correctly.

Surface of objective lens is 
dirty. 

Clean it

Immersion oil isn’t used for 
100

×

 objective (oil).

Use immersion oil

There is a bubble in the 
immersion oil.

Clear the bubble away

Size of aperture

 

diaphragm 

isn’t correct.

Adjust the size of 
aperture diaphragm .

Position of condenser is too 
low.

Rotate condenser up.

- - 

6

Содержание VM200 Series

Страница 1: ...VM200 Series Biological Microscope Instruction Manual To ensure your safety and to obtain satisfactory performance please study this manual before using the microscope VM200...

Страница 2: ...mm 75mm 5 Focusing system Coaxial Coarse and Fine Focusing Knobs Coarse stroke maximum 25mm Fine division 2 m 6 Condenser Abbe N A 1 25 Adjustable aperture up down range 12mm 7 LED Light Source Input...

Страница 3: ...tage A5 Abbe Condenser A6 Illuminator A7 Base A8 Main Body A9 DC Input Socket A10 Power Adapter A11 Power Cable Operational Parts B1 Nosepiece B2 Slide Holder B3 Iris Diaphragm B4 Tension Adjusting Ri...

Страница 4: ...e it is recommended that it be serviced regularly In case of damage contact Valley Microscope via email service valleymicroscope com Any unauthorized service repairs may void the warranty 4 Observing...

Страница 5: ...trument will be seriously damaged 3 Preventing burns and fire Alcohol gasoline paper and other flammable materials must be kept away from the illuminator 4 Carrying and operation Power must be turned...

Страница 6: ...position The size of aperture diaphragm is too small Adjust the size of aperture diaphragm Lens objective eyepiece condenser illuminator collector has dust Clean lens Position of condenser is too low...

Страница 7: ...level in specimen holder Objective touches specimen when changing from low power to higher power Cover glass of specimen doesn t meet the requirement Place specimen correctly Cover glass is too thick...

Отзывы: