10)
Connect the 2300 as shown in Figure
8-4 using the 10 amp shunt.
11)
Select the 10 amp range.
12)
Apply 10.000 amps and adjust
φ
AR5
for 5.000 VDC on the DVM.
13)
Connect the 2300 as shown in Figure
8-4 using the 100 amp shunt.
14)
Select the 100 amp range.
15)
Apply 100.00 amps and adjust
φ
AR7
for 5.000 VDC on the DVM.
16)
Disconnect the DVM and current
calibrator.
8-5-4. Power Calibration
1) Connect the 2300 as shown in Figure 8-7.
2) Select
φ
A and the 150 (15) volt and 1 amp
ranges.
3) Connect the DVM low lead to the
negative leg of MBC5 and the high lead to
the front leg of MBR9.
4) Apply 150.00 volts and 1.0000 amps and
adjust
φ
AR40 for 2.500 VDC.
5) Disconnect the DVM and calibrators.
8-6.
Option IOX Calibration
Option IOX should be calibrated following
successful completion of the Routine
Calibration procedure of Section 8-4. In this
procedure, all measurements are made at the
IOX connector using a DC voltmeter. All
measurements are made with DVM high lead.
The DVM low lead should be connected to
the IOX connector common at pin 12.
Adjustments are accessed by removing the
bottom cover of the instrument.
Caution!
AC line power is exposed at this point. Take
precautions to avoid contact with AC power
wiring.
Note:
In simultaneous TL-4/IOX
installations, adjustments are accessed from
the top of the instrument and removal of the
bottom cover is not necessary.
8-6-1. Channel A IOX Calibration
1) Connect the 2300 as shown in Figure 8-7.
Select
φ
A, 150 (15) volts, 1 amp ranges.
2) Apply 150.00 volts and adjust RV12 for
an output of 5.000 volts at pin 1 of the
IOX connector.
3) Apply 15.00 volts and adjust RV1 for an
output of .5000 volts at pin 1 of the IOX
connector.
4) Repeat steps 2 and 3 for minimum
interaction.
5) Apply 1.0000 amps and adjust RV13 for
an output of 5.000 volts at pin 2 of the
IOX connector.
6) Apply .1000 amps and adjust RV2 for an
output of .5000 volts at pin 2 of the IOX
connector.
7) Repeat steps 5 and 6 for minimum
interaction.
8) Apply 150.00 volts and 1.0000 amps;
adjust RV14 for 5.000 volts at pin 3 of the
IOX connector.
9) Apply 15.00 volts and 1.0000 amps and
adjust RV7 for .5000 volts at pin 3 of the
IOX connector.
10) Repeat steps 8 and 9 for minimum
interaction.
11)
Apply 150.00 volts and 1.0000 amps;
adjust RV15 for 2.500 volts at pin 10
of the IOX connector.
12) Apply 15.00 volts and 1.0000 amps
and adjust RV10 for .2500 volts at pin
10 of the IOX connector.
13)
Repeat steps 11 and 12 for minimum
interaction.
14)
Apply 150.00 volts and
1.0000 amps and adjust
Содержание 2300
Страница 3: ...This manual covers the following Valhalla Scientific products Models 2300 2301 2300L and 2301L...
Страница 8: ...SECTION XII MANUAL CHANGES AND ADDENDUMS SECTION XIII PARTS LISTS SECTION XIV DRAWINGS AND SCHEMATICS...
Страница 19: ......
Страница 20: ......
Страница 23: ......
Страница 24: ......
Страница 26: ......
Страница 33: ......
Страница 34: ......
Страница 35: ......
Страница 36: ......
Страница 37: ......
Страница 38: ......
Страница 39: ......
Страница 40: ......
Страница 41: ......
Страница 42: ......
Страница 43: ......
Страница 44: ......
Страница 50: ......
Страница 64: ......
Страница 65: ......
Страница 66: ......
Страница 67: ......
Страница 68: ......
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 74: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 95: ......
Страница 96: ......
Страница 97: ......
Страница 98: ......
Страница 100: ......
Страница 101: ......
Страница 103: ......
Страница 106: ......
Страница 108: ......
Страница 109: ......
Страница 110: ......
Страница 111: ......
Страница 113: ......
Страница 114: ......
Страница 115: ......