ProLogger Operational Status
Solid Green
Run mode
Flashing Green
Interrupted mode
Amber
Bootloading Mode
(only lit when upgrading firmware)
Bootloader Reset
In order to Bootload new firmware into the ProLogger the unit has to be reset. In order to do this the Magnetic
On/Off key must be placed close to the white, translucent end cap and pins 7 and 8 shorted together. Any
Bootloading process must be carried out serially (NOT through Bluetooth) using the USB interface cable direct to a
PC.
Outside of a rapidPro the ProLogger is not powered. To power the ProLogger outside of the instrument it must be
powered with the charging/data cable provided using either the charger provided or a PC USB port.
The ProLogger is compatible with both a rapidPro SVT and rapidPro CTD and can be swapped between the two.
2.3.1. Changing\Charging the ProLogger
·
Changing\Charging the ProLogger of a rapidPro is a very simple process. All data is saved in non-volatile
Flash memory, so data is secure during the process
·
Remove the grub screw just below the black rear housing of the RapidPro profiler and then carefully unscrew
rear housing
-
An Allen Key that fits the grub screw is provided in the case
·
The ProLogger can now be carefully removed for charging/data download
-
charging is carried out using the provided cable and charger
-
Data download operations can be carried out using the same charging cable as above but interfaced to a
PC's USB port
The ProLogger must be correctly 'ejected' from the PC
·
Check for any water ingress to the instrument that has occurred during this process. Dry if necessary
Before replacing the ProLogger into the rapidPro all LEDs should be off
If the LEDs are still on, re-interface the unit to the PC and 'eject' correctly
·
Replace the ProLogger into the instrument - rotating slowly until the location pins align, the ProLogger will
drop further into the rapidPro indicating that it is correctly sited
·
Check the o-ring for damage and debris. Clean it and smear lightly with silicon grease. Replace if
necessary
For planned periods of extended storage, the ProLogger should be removed from the instrument
Содержание rapid ProSVT
Страница 5: ... 2019 Valeport Ltd EU Declaration of Conformity CE Marking Page 5 ...
Страница 9: ... 2019 Valeport Ltd Introduction Page 9 2 1 2 Dimensions ...
Страница 17: ... 2019 Valeport Ltd Introduction Page 17 2 2 6 Dimensions rapidPro CTD and rapidPro CTDplus ...
Страница 34: ... 2019 Valeport Ltd 0660907a rapidPro Operating Manual Page 34 ...