miniIPS2 and 2e Operating Manual: MANUAL-383434403-1
Page | 21
©
2020
–
Valeport Ltd
6.5
Format: Valeport NMEA ($PVIPS2)
Command:
#013;8
Format:
$PVIPS2,aa,bbbbb,p.ppp,d,d.ddd,y*zz
Notes:
NMEA identifier,instrument address, serial number,pressure in dBar, d, Data in units
(#018),unit Checksum
Field
Format
NMEA Identifier
Instrument address
aa
Serial number
bbbbb
Pressure
p.ppp
Unit
d (dBar)
Depth
d.ddd
Depth type
y
Checksum
*zz
The units displayed in the string can be enabled | disabled
Output String Examples
With units
$PVIPS2,01,45687,10.996,d,10.988,M*61
$PVIPS2,01,45687,10.996,d,10.988,M*61
Without units
$PVIPS2,01,45687,10.996,d,10.988,*2c
$PVIPS2,01,45687,10.996,d,10.988,*2c
6.6
Format: CSV
Command:
#013;1
Format:
dd/mm/yy, hh:mm:ss, xxxx.x,0.0,00.0<CR><LF>
Notes:
Field
Format
Date
DD/MM/YY
Time
HH:MM:SS
Pressure or Depth
xxxx.x
Spare
0.0
Spare
0.00
The miniIPS2 does not have a real time clock so the Time and Date fields are not
populated. The fields are maintained for compatibility purposes
Output String Example
00/00/00, 00:00:00, 0010.0,0.0,0.00
00/00/00, 00:00:00, 0010.1,0.0,0.00