5.2 Command Codes
The fastCTD is designed to be used with Valeport’s DataLog X2 software. However, the
instrument will respond to a series of text commands that are detailed here. Note that this list is not
comprehensive, but will allow the standard functions of the instrument to be accessed. For more
detailed information, please contact Valeport Ltd.
Notes
All commands must be confirmed using “Carriage Return” or “Enter” on the keyboard, with the
exception of the “Stop” command (#).
All commands are echoed back by the instrument as they are typed
5.2.1 Stop Command
The instrument can be stopped at any time by sending the ‘#’ character. The instrument returns a
‘>’, and waits for a further command.
5.2.2 Run Commands
R1
Collects rapid profile at 1 Hz
R2
Collects rapid profile at 2 Hz
R4
Collects rapid profile at 4 Hz
R8
Collects rapid profile at 8 Hz
R16
Collects rapid profile at 16 Hz
R32
Collects rapid profile at 32 Hz
M1
Performs continuous measurement at 1Hz
M2
Performs continuous measurement at 2Hz
M4
Performs continuous measurement at 4Hz
M8
Performs continuous measurement at 8Hz
M16
Performs continuous measurement at 16 Hz
M32
Performs continuous measurement at 32 Hz
Running at high update rates will provide increased temporal sensitivity but will result in noisier
data, especially optical.
Содержание fastCTD
Страница 11: ...fastCTD and fastCTD plus MANUAL 2034990662 4 issue 1 0 Page 9 2021 Valeport Ltd 4 1 Dimensions...
Страница 29: ...fastCTD and fastCTD plus MANUAL 2034990662 4 issue 1 0 Page 27 2021 Valeport Ltd 13 1FastCTD and fastCTD plus...
Страница 30: ...EU Declaration of Conformity CE Marking 2021 Valeport Ltd Page 28 13 2fastCTD with Blutooth Option...