Model 001 & 002 Flowmeter Operating Manual
MANUAL-1550859975-1 | issue: 1.1
©
2021
– Valeport Ltd
Page | 15
5.9.1 General Control Codes #000 - #014
Code
Followed By
space and
Operation
Access
#000
Exit calibration set-up mode
All
#001
New calibration
string<cr>
OTHER 1 IMPELLER
Changes calibration to New_calibration string sent must
be to the required format using spaces as separators
All
#002
New calibration
string<cr>
OTHER 2 IMPELLER
Changes calibration to New_calibration string sent must
be to the required format using spaces as separators
All
#003
New calibration
string<cr>
OTHER 3 IMPELLER
Changes calibration to New_calibration string sent must
be to the required format using spaces as separators
All
#004
New calibration
string<cr>
OTHER 4 IMPELLER
Changes calibration to New_calibration string sent must
be to the required format using spaces as separators
All
#011
New name
string<cr>
OTHER 1 IMPELLER
Changes the name of other 1 impeller from other 1 to new
name string - for ease of user identification.
All
#012
New name
string<cr>
OTHER 2 IMPELLER
Changes the name of other 2 impeller from other 2 to new
name string - for ease of user identification.
All
#013
New name
string<cr>
OTHER 3 IMPELLER
Changes the name of other 3 impeller from other 3 to new
name string - for ease of user identification.
All
#014
New name
string<cr>
OTHER 4 IMPELLER
Changes the name of other 4 impeller from other 4 to new
name string - for ease of user identification.
All
5.9.2 Table 1: Calibration Function Numbers and Meaning
Calibration Function No.
OPERATION
DEFINED
0
Not defined
NO
1
Straight line fit
YES
2
Polynomial fit
YES
5.9.3 How to Program Calibration Coefficients.
First, decide if your calibration is a polynomial equation or a series of straight line fits. For line fits,
proceed to Section:
. For polynomial fits, proceed to Section: