Bioscanner Manual
www.vaki.is
Page 22 of 25
5.
T
ROUBLESHOOTING
5.1.
O
VER COUNTING
If the water flow in the pipeline is uneven and the water is splashing inside the scanner, try to
reduce the water or adjust the water flow. Check and ensure that the fish not going backwards
through the counter again. Check if the fish size is correctly set on the Control Unit.
5.2.
U
NDERCOUNTING
Check if the fish size is correctly set on the Control Unit. When too many fish are pumped
through the Scanner Unit, the counter is not able to count all the fish because the images are
overlaying each other. Check the rate of fish to ensure the counter is not overloaded. Try to
reduce the amount of fish, and make sure the fish are not blocking up in the pipeline but gliding
smoothly through the pipe. If there is not enough water for the fish to slide smoothly down the
pipe, this can also cause a blockage.