4.1.2 Setting serial communication parameters
The following parameter combinations are available for PTB210. The factory setting for the
parameters is E71.
.E71
or
.O71
or
.N81
The command sets the serial communication parameters of either a single barometer or all the
barometers.
4.1.3 .ID setting identification code
.ID.xxx
xxx
Identification code (max. 15 characters)
The default ID code is 0. When several barometers are on the same line, use the ID
code to direct the command to a specific instrument separately.
The barometer handles the ID code as characters, for example 0001, 01 and 1 are
different codes
4.1.4 .PMIN / .PMAX setting pressure limits
You can define a lower and a higher pressure limit to the barometer. You can define, for
example, the calibrated pressure range to a host system.
.PMIN.xxxx
xxxx
Lower pressure limit (0 ... 15000 hPa)
If current pressure is under the limit, the barometer outputs ****.** instead of the pressure
reading.
.PMAX.xxxx
Chapter 4 – Operation
15