User's Guide ______________________________________________________________________
164 _________________________________________________________________ M210667EN-C
Status Codes
LT31 System
For decoding the status, see Table 13 below.
Table 13
Status Codes
Byte Number
Description
21
LT31 overall status
22
LT31 MOR status
23
LT31 power supply
24 LT31
auto-calibration
25 LT31
window
contamination
26 LT31
alignment
27
LT31 reserved for future extension
28
LT31 reserved for future extension
29 LT31
transmitter
status
30
LT31 transmitter -> main transmitter module
31
LT31 transmitter -> window transmitter module
32
LT31 transmitter -> measurement CPU
33
LT31 transmitter -> misc.
34 LT31
receiver
status
35
LT31 receiver -> master CPU status
36
LT31 receiver -> main receiver module
37
LT31 receiver -> window transmitter module
38
LT31 receiver -> measurement CPU
39
LT31 receiver -> misc.
40 PWD
status
For decoding the overall status, see Table 14 below.
Table 14
LT31 Overall Status (Byte Number 21)
Code Explanation Comment
0 OK
I Indication
of
abnormal
situation
AC off, operating on battery.
Auto-alignment in process.
W Warning Examples:
Window contamination reached
warning level.
Transmitter LED aging.
Backup battery voltage low.
A Alarm
Examples:
Communication error in LT31 system.
Internal monitoring alarm signal
exceeded limit.
E Error
Examples:
Transmitter, receiver, PWD (if
configured) or LM21 (if configured) is
missing.
S
Service
User interface has been opened.