5. Startup
5.1 Startup
5.1.1 Initial check
Connect the refractometer, and check that it powers up properly, see
. Use a web browser to open the refractometer homepage and check that
the serial number of the refractometer corresponds to that on the refractometer nameplate.
Check the parameters on the
Parameters
page and adjust them as needed, see
. Remember to submit changes.
5.1.2 mA check
Measure the mA output. If the refractometer is not in the line (i.e. message is
NO SAMPLE
), mA
out sends default mA. The default mA and mA maximum and minimum are set via web
browser on the
Parameters
page, see
5.1.3 Calibration check
Wait until normal process conditions occur. The concentration reading is precalibrated at
delivery and a copy of the Refractometer calibration certificate is shipped with the
refractometer. If the diagnostic message is
Normal operation
but the concentration reading
does not agree with the laboratory results, see
Calibrating concentration measurement
5.2 Viewing refractometer status
The basic information on the measurements is shown on the
Main
page of the refractometer,
. More information is shown on the
Diagnostics
page, see
The measurement result is calculated from the refractive index (n
D
and process temperature
(T) values. Both of these values are on the
Main
page.
In addition to these measurements, the refractometer monitors its internal temperature and
humidity, which both are available on the
Diagnostics
page. The internal temperature should
not be above 55 °C, and the humidity should be below 30 %.
Chapter 5 – Startup
23
Содержание K-PATENTS PR-43
Страница 1: ...M212455EN B User Guide Process Refractometer PR 43 Series...
Страница 24: ...Figure 18 Verification page PR 43 Series User Guide M212455EN B 22...
Страница 50: ......
Страница 52: ......
Страница 53: ......
Страница 54: ...www vaisala com...