10
PR-43-...-IA/AX/CU instruction manual
Document/Revision No.
IM-EN-PR43IAAX
/A
Effective: October 1, 2019
24VDC
Refractometer
PR-50013 Isolator
PR-10930 P4 ISOLATOR
+ -
Green
White/Green
Refractometer IS cable PR-8436
24VDC
- Brown
+ White/Brown
Cable supported to enclosure
COVER
>5 mm²
Ex Ground
9
8
14
7
6
5
4
3
No connetion
Ethernet
Connector cover can be opened only with tool
Tape back shield
Refractometer shall be
Ex-grounded through the
process piping. In case of
nonconductive piping
(e.g. plastic or glass)
refractometer shall be
grounded one of the
thread holes in sensor cover.
10
13
12
11
White/Orange Orange
PR-8432 Power supply and mA output cable
+
-
mA Output 1 4-20mA
Cable supported to enclosure
Eth RX-
Eth TX-
Eth RX+
Eth TX+
No connection
2mm2
Ex ground
Ex ground
Ex ground Grounding nut M5
Tape back shield
mA
mA output -
2
1
2
Figure 3.5
Wiring PR-43-...-IA with isolator
and PR-8432 cable for mA output and power
Содержание K-PATENTS PR-43-AX Series
Страница 1: ...IM EN PR43IAAX A Instruction Manual Vaisala K PATENTS Process Refractometer PR 43 IA AX CU...
Страница 4: ...Document Revision No IM EN PR43IAAX A Effective October 1 2019...
Страница 6: ...PR 43 IA AX CU instruction manual Document Revision No IM EN PR43IAAX A Effective October 1 2019...
Страница 32: ...26 PR 43 IA AX CU instruction manual Document Revision No IM EN PR43IAAX A Effective October 1 2019...
Страница 34: ...28 PR 43 IA AX CU instruction manual Document Revision No IM EN PR43IAAX A Effective October 1 2019...
Страница 35: ......
Страница 36: ...www vaisala com...