User's Guide ______________________________________________________________________
164 _________________________________________________________________ M210566EN-J
Feeding Adjustment Information
This information is shown on the device information fields. See section
Device Information on page 107.
Using Display/Keypad
1.
If you are not in the adjustment menu, press the
ADJ
button on the
motherboard (opens the
ADJUSTMENT MENU
).
2. Select
Adjustment info,
press the
►
button.
3. Select
Date
, press
SET
. Give the date by using the arrow buttons.
Press
OK
.
4. Select
i,
press
SET.
Enter information text including 17 characters
at maximum by using the arrow buttons. Press
OK.
5. Press
EXIT
to return to the basic display.
Using Serial Line
CTEXT
Use the CTEXT command to enter text to the adjustment information
field.
Example:
>
ctext
Adjust. info : (not set) ?
HMK15
>
CDATE
Use the CDATE command to enter date to adjustment information field.
Set the adjustment date in format YYYY-MM-DD
.
Example:
>
cdate
Adjust. date : (not set) ?
2004-05-21
>