5.2 Errors
In an error state, HMP155 does not measure the quantity and shows the following output:
• Analog channel:
0 V
Use the
AERR
serial line command to change the value.
• The serial port output:
***
You can check the error message through the serial interface by using the
ERRS
command. If
the error persists, contact Vaisala Technical Support.
If you do not know the HMP155 serial connection settings, you can force the
settings 19200 N 8 1 with the
#
command. The
#
command is only available
during the first 3 seconds after startup.
Table 29 HMP155 error messages
Error message
Description
Action
T MEAS error
Error in temperature
measurement
Check the HUMICAP sensor.
T REF error
Contact Vaisala Service Center.
TA MEAS error
Error in T-probe
measurement
Check the additional temperature
probe.
TA REF error
Contact Vaisala Service Center.
F MEAS error
Error in humidity
measurement
Check the HUMICAP sensor.
F REF1 error
Contact Vaisala Service Center.
F REF3 error
Program flash
checksum error
Internal error
Parameter flash
checksum error
INFOA checksum error
SCOEFS checksum error
More information
‣
Chapter 5 – Maintenance
65
Содержание HUMICAP HMP155
Страница 1: ...M210912EN D User Guide Vaisala HUMICAP Humidity and Temperature Probe HMP155 ...
Страница 8: ...HMP155 User Guide M210912EN D 6 ...
Страница 90: ...7 7 3 Radiation shield dimensions Figure 17 DTR503 and DTR13 dimensions HMP155 User Guide M210912EN D 88 ...
Страница 94: ......
Страница 98: ...HMP155 User Guide M210912EN D 96 ...
Страница 100: ......
Страница 101: ......
Страница 102: ...www vaisala com ...