Example
To set the scaling of the outputs without changing the quantities, type the command without
parameters. HMP155 prompts for the limits. Note that
ch0
is channel 1 and
ch1
is channel 2.
>asel
Ch0 T lo : -40.00 'C ?
Ch0 T hi : 60.00 'C ?
Ch1 RH lo : 0.00 %RH ?
Ch1 RH hi : 100.00 %RH ?
Example
To set the output quantities of the channels, type the quantities as parameters. The probe
prompts for the limits. Always specify the quantities for both channels.
>asel t td
Ch0 T lo : -40.00 'C ?
Ch0 T hi : 60.00 'C ?
Ch1 Td lo : -60.00 'C ?
Ch1 Td hi : 40.00 'C ?
Example
To set the quantities and scaling with a single command, type the quantities and limits as
parameters according to the command syntax.
>asel rh td 0 100 -60 60
Ch0 RH lo : 0.00 %RH
Ch0 RH hi : 100.00 %RH
Ch1 Td lo : -60.00 'C
Ch1 Td hi : 60.00 'C
4.8.4 AERR command
Use the
AERR
serial line command to set or view the error output from analog outputs.
AERR
CH0ERR/CH1ERR
Parameter
Description
CH0ERR
Analog output
CH1ERR
Analog output
Chapter 4 – Operation
55