USER'S GUIDE____________________________________________________________________
SEND Outputting a Reading Once
SEND <cr> in STOP state
or
SEND aa <cr> in POLL state
aa =
address of the transmitter when more than one transmitter
is connected to a serial bus (0 ... 99; set with command
ADDR)
Outputs the current measurement readings via the serial line. The output
format depends on which parameters the transmitter can output. Output
types are:
"RH 32.25 %RH T 25.74 'C ",<cr><lf>
"Td 7.93 'C",<cr><lf>
INTV Setting the Output Interval for the
RUN State
INTV xxx yyy <cr>
xxx =
output interval (0 ... 255)
0: no pause between outputs
yyy =
unit (s, min or h)
Sets the output interval when the transmitter outputs measurement
readings to a peripheral device.
For example:
>INTV 10 MIN<cr>
Output intrv. : 10 min
22 ____________________________________________________________________ U339EN-1.3