Register
number
Address
Register description
Data
format
Note
520
0207
hex
RH measurement status
16-bit
signed
integer
0000
hex
: Status OK
0001
hex
: Measurement is not
available
0002
hex
: Measurement is not
reliable
0004
hex
: Under range
0008
hex
: Over range
0020
hex
: Value locked
0080
hex
: Sensor failure
0100
hex
: Measurement is not
ready
521
0208
hex
T measurement status
16-bit
signed
integer
522
0209
hex
Tdf measurement status
16-bit
signed
integer
523
020A
hex
Device status
16-bit
signed
integer
0000
hex
: Status OK
0001
hex
: Critical error,
maintenance needed
0002
hex
: Error, device may
recover automatically
0004
hex
: Warning
0008
hex
: Notification
0010
hex
: Calibration mode
active
Table 22 Error codes in register 0203
hex
Bitmask
Error message
Severity
0000
hex
Status OK
0001
hex
Temperature measurement error
Error
0002
hex
Humidity measurement error
Error
0004
hex
Humidity sensor failure
Error
0008
hex
Capacitance reference error
Error
0010
hex
Ambient temperature out of range Error
0800
hex
Supply voltage out of range
Error
800000
hex
Sensor heater failure
Warning