Address
(Decimal)
Address
(Hexadecimal)
Register Description
Data
Format
Notes
529
0210
hex
Status of Hydrogen
peroxide concentration
by volume
16-bit
0 = Status OK.
2 = Reading is not
reliable.
32 = Reading is locked
during purge.
128 = Sensor failure.
256 = Measurement not
ready
530
0211
hex
Status of H
2
O+H
2
O
2
relative saturation
16-bit
531
0212
hex
Status of Temperature
16-bit
532
0213
hex
Status of Relative
humidity
16-bit
533
0214
hex
Status of Absolute
hydrogen peroxide
16-bit
536
0217
hex
Status of Water
concentration by volume
16-bit
537
0218
hex
Status of Water vapor
pressure
16-bit
538
0219
hex
Status of Absolute
humidity
16-bit
539
021A
hex
Status of Water vapor
saturation pressure
16-bit
540
021B
hex
Status of Temperature
from the temperature
probe
16-bit
1)
Multiple statuses can be present simultaneously. In those cases, the value of the status
register is the sum of the status values. For example the value of the device status register is
6
if a warning (
4
) and an error (
2
) are present simultaneously.
Table 16 Error Codes in Register 0201
hex
(32-bit)
Value of
Register
0201
hex
Error Message
Severity
0
Status OK.
1
Firmware checksum mismatch.
Critical
2
Device settings corrupted.
Critical
8
Supply voltage too high.
Error
16
Internal voltage too low.
Error
32
Internal voltage too high.
Error
64
Humidity sensor failure (open circuit).
Error
Appendix A – Modbus Reference
55