Address
(Decimal)
Address
(Hexadecimal)
Register Description
Data
Format
Unit / Valid Range
1793
0700
hex
Output mode
enum
0 = Off
1 = 0 ... 20 mA
2 = 4 ... 20 mA
(default: 2 = 4 ... 20 mA)
1794
0701
hex
Output parameter
16-bit
register
address
Valid range
0000
hex
... 0016
hex
.
For the available
parameters and their
register addresses, see
.
1795
0702
hex
Scale low end
32-bit float Set the lower limit of the
measurement scale for
the output parameter
chosen in register 0701
hex
(Output parameter).
1797
0704
hex
Scale high end
32-bit float Set the upper limit of the
measurement scale for
the output parameter
chosen in register 0701
hex
(Output parameter).
1799
0706
hex
Error output
32-bit float mA
0 ... 25
Analog Output 2:
2049
0800
hex
Output mode
enum
0 = Off
1 = 0 ... 20 mA
2 = 4 ... 20 mA
(default: 2 = 4 ... 20 mA)
2050
0801
hex
Output parameter
16-bit
register
address
Valid range
0000
hex
... 0016
hex
.
For the available
parameters and their
register addresses, see
.
2051
0802
hex
Scale low end
32-bit float Set the lower limit of the
measurement scale for
the output parameter
chosen in register 0801
hex
(Output parameter).
HPP272 User Guide
M211972EN-B
52