10. Technical data
10.1 HMP80 specifications
Table 6 HMP80 series measurement performance
Property
Description/Value
Relative humidity
Measurement range
0 … 100 %RH
Accuracy at +23 °C (73.4 °F)
±0.8 %RH (0 … 90 %RH)
Factory calibration uncertainty
±0.5 %RH (0 … 40 %RH)
±0.8 %RH (40 … 95 %RH)
T
63
response time
15 s
Sensor
HUMICAP
â
R2C
Temperature
Measurement range
HMP80N: −20 … +60 °C (−4 … +140 °F)
HMP80L: −50 … +120 °C (−58 … +248 °F), short-
time measurement range
−50 … +180 °C (−58 … +356 °F)
Accuracy at +23 °C (+73.4 °F)
±0.1 °C (±0.18 °F)
Factory calibration uncertainty
±0.1 °C (±0.18 °F) at +23 °C (+73.4 °F)
Sensor
Pt100 RTD Class F0.1 IEC 60751
1)
Defined against calibration reference. Including non-linearity, hysteresis, and repeatability.
2) Defined as ±2 standard deviation limits. Small variations possible; see calibration certificate.
3) Exposing temperature sensor to temperatures below −20 °C (−4 °F) may cause permanent
additional deviation of ±0.1 °C (0.18 °F).