3. To adjust temperature measurement:
a. If you did not adjust humidity measurement, select the
Calibration
tab and select
Start calibration
to start the calibration mode.
b. In the
T adjustment
section of the page, select
Restore factory adjustment
to remove
any existing non-factory adjustments.
c. Insert the probe head in the reference environment for T adjustment point 1.
d. Wait for temperature measurement to stabilize fully. This may take more than 30
minutes. Monitor the reading to see when the measurement has stabilized.
e. Enter the value of the first reference into the
Reference value, point 1
field. Select
outside the text box when done. The probe automatically enters the measured value
for the adjustment point.
f. If you want to adjust temperature measurement in two points, repeat steps
for T adjustment point 2.
g. Select
Activate adjustment
to store the adjustment in the probe.
4. In the
Calibration information
section of the page, update the
Calibration date
and
Calibration text
fields.
5. Select
Stop calibration
to end the calibration mode.
Chapter 6 – Maintenance
45