User's Guide _______________________________________________________________________
46 ___________________________________________________________________ M211060EN-H
filt
Filter : 1.000 ?
Example (set filtering to 0.5):
filt 0.5
Filter : 0.500
Set Probe Address
Use the
ADDR
command to view or set the probe address. To operate in
the POLL mode, the probe must have an address. If multiple probes share
the same serial line, each probe must have a different address.
For a description of the serial interface modes, see section Set Serial Line
ADDR
[
nn
]<cr>
where
nn
= address (0 ... 255)
Example:
addr
Address : 0
Set Serial Interface Delay
Use the
SDELAY
command to view or set the serial interface answer
minimum delay.
SDELAY
[
delay
]<cr>
where
delay
= Range 0 ... 255. Value corresponds to four
milliseconds (for example, 5 = 0.020 second
minimum answer delay)
Example:
sdelay
Serial delay : 30
sdelay 50
Serial delay : 50