Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 43
Outputting the results continues in intervals issued with the command
INTV
. You can stop the output by entering the
S
command.
Stop Measurement Output
Use the
S
command or press the
Esc
key to stop the continuous
measurement output.
S
<cr>
Output the Measurement Message Once
Use the
SEND
command to output the measurement values once. If the
probe is in POLL mode and the line is not open for commands, specify
the address of the probe to receive the measurement message.
SEND
[
aaa
]<cr>
where
aaa = Address of the probe, range 0 ... 255. Set with the
ADDR
command.
Example (probe in STOP mode, no address needed):
send
T= 22.7 'C RH= 20.0 %RH Td= -1.5 'C
Example (probe in POLL mode, with address 10):
send 10
T= 22.8 'C RH= 20.1 %RH Td= -1.3 'C
Configuring Serial Line Operation
Set Serial Line Settings
Use the
SERI
command to show or set the serial line settings. The new
settings will be taken into use when the probe is reset or powered up.
SERI
[
b p d s
]<cr>
where
b = baud rate
(300, 600, 1200, 2400, 4800, 9600, 19200, 38400, 57600)
p = parity (n = none, e = even, o = odd)
d = data bits (7 or 8)
s = stop bits (1 or 2)