HMM30C MODULE

HMM30C

U312en-1.3

Operating Manual

99-11-29

The calibration interval depends on the surrounding conditions and the desired accuracy.
It is recommended to perform a calibration at least once a year. However, long term
exposure to certain chemicals and gases, e.g. some organic solvents, acids or bases may
cause drift in the sensor response. If there is a reason to believe that such conditions exist,
it is recommended to check the calibration at a shorter interval after taking the HMM30C
into use. Make the first check/adjustment after 6 months and 12 months and later extend
the interval to one year.

REPLACEMENT OF THE HUMICAP

 SENSOR OR THE FILTER

If the sensor is damaged for some reason, remove it and insert a new one. Handle the
sensor by the plastic socket. DO NOT TOUCH THE SENSOR PLATE. Check the output
after sensor change. If it deviates too much from the reference value, recalibrate the
module.

Replace or clean a dirty filter. The sensor can be cleaned with distilled water; if this does
not help, replace the sensor. Do not use any mechanical methods.

TECHNICAL DATA

Relative humidity

Measurement range

0...100 %RH

Accuracy at +20 °C
(including calibration
inaccuracy, linearity and
repeatability):

±2 %RH (0...90 %RH)
±3 %RH (90...100 %RH)

Temperature coefficient

±0.04 %RH/°C (NOTE: after T compensation)

Humidity sensor

Vaisala HUMICAP

 capacitive polymer sensor

Electronics

Output

Supply voltage

DC

DC

AC

0...1 V

10...35 V

9...24 V

0...5 V

14...35 V

12...24 V

0...10 V

19...35 V

16...24 V

0...20 mA

10...35 V

9...24 V (R

L

 = 0 ohm)

0...20 mA

20...35 V

17...24 V(R

L

 = 500 ohm)

Recommended external
load for outputs

> 10 kohm

Current consumption

15 mA max.

Содержание HMM30C

Страница 1: ...two screws Its small sensor head is separated from the circuit board with a cable Mounting dimensions are shown in Figure 1 215 8 46 ø12 1500 59 05 49 1 93 24 0 94 55 2 16 0 47 42 5 1 67 Figure 1 Mounting dimensions in mm in inches ELECTRICAL CONNECTIONS Make connections according to Figure 2 OUT 2 1 3 DC AC POWER SUPPLY V mA Figure 2 Electrical connections ...

Страница 2: ...output range is chosen with appropriate jumpers see Figure 3 When the output has been chosen and the module has been connected to an external power supply it is ready for use RQH SRLQW FDOLEUDWLRQ FRQQHFWRU RXWSXW VHOHFWLRQ 9 9 9 P GU HQG FDOLEUDWLRQ ZHW HQG FDOLEUDWLRQ Figure 3 Output selections and calibration connectors and potentiometers ONE POINT HUMIDITY CALIBRATION The calibration can be ch...

Страница 3: ...remove it and insert a new one Handle the sensor by the plastic socket DO NOT TOUCH THE SENSOR PLATE Check the output after sensor change If it deviates too much from the reference value recalibrate the module Replace or clean a dirty filter The sensor can be cleaned with distilled water if this does not help replace the sensor Do not use any mechanical methods TECHNICAL DATA Relative humidity Mea...

Страница 4: ...ics 0 85 RH Mechanics Sensor protection Sintered filter Ø 12 mm PCB dimensions 55 x 42 5 x 23 mm Weight 144 g Cable length 1 5 m 5 ft SPARE PARTS AND ACCESSORIES Order code Description 0195 sintered filter 12 mm HUMICAP K HUMICAP humidity sensor GUARANTEE Vaisala issues a guarantee for the material and workmanship of this product under normal operating conditions for one 1 year from the date of de...

Отзывы: