Syntax
Description
Example: set output format as RH and T, with start of text (ASCII character 002) and end of text
(003) ASCII codes, and without line feed and carriage return at the end:
form #002 3.1 "RH=" RH U4 3.1 "T=" T " " U3 #003
OK
Output example (ASCII codes not visible here):
send
RH= 39.3%RH T= 25.1 'C
Table 18 Output Parameters for Form Command
Measured
Parameter
Abbreviation in Form Command
Temperature
t
Relative humidity rh
Dew point
td
Enthalpy
h
Wet bulb
temperature
tw
Table 19 Modifiers for Form Command
Modifier
Description
x.y
Length modifier (number of digits and decimal places)
\t
Tabulator
\r
Carriage return
\n
Line feed
\xxx
ASCII character with decimal code xxx
""
String constant
ux
Name of the measurement unit using x number of characters. For example, u3
shows the name of the measurement unit with three characters.
addr
Transmitter address
sn
Transmitter serial number
time
Time since transmitter was started or reset
HMDW110 User Guide
M211726EN-D
42
Содержание hmdw110 series
Страница 1: ...M211726EN D User Guide Vaisala Humidity and Temperature Transmitter Series HMDW110 ...
Страница 5: ...Recycling 87 Table of Contents 3 ...
Страница 14: ...2 5 HMS110 112 Parts 1 2 3 4 5 6 7 8 9 10 11 Figure 3 HMS110 112 Parts HMDW110 User Guide M211726EN D 12 ...
Страница 25: ...3 Mount the transmitter using two screws of sufficient length 100 mm Ø 5 5 mm Chapter 3 Installation 23 ...
Страница 87: ...Appendix A Modbus Reference 85 ...
Страница 88: ...HMDW110 User Guide M211726EN D 86 ...
Страница 90: ...HMDW110 User Guide M211726EN D 88 ...
Страница 91: ......
Страница 92: ...www vaisala com ...