HMI41 Calibration Option

19165ZZ-U220en-1.1

Appendix 1

1996-08-26

1

APPENDIX 1: QUICK REFERENCE TO CALIBRATION

It is recommended that this quick reference guide is used only as a checking
list for those who already know how to operate the HMI41 as a calibrator. For
those who take it into use for the first time, the Operating Manual gives useful
information that is not included in this quick reference guide.

1.

Selecting the calibrator function

For selecting the calibrator function, press ON/OFF until you can see some
text on the display. Then release the ON/OFF button and press within 1...2
seconds both ENTER and MODE buttons until the text "setup" appears on the
display. Then follow the instructions of the table below.

DISPLAY

WHAT TO DO

PRESS:

Wait for a few seconds.

set

°C

If the basic settings have been given,
press ENTER.
If they have not been given, refer to the
HMI41 Operating Manual.

press ENTER repeatedly:

set

select the HMI41 function:
1 = indicator
2 = calibrator for HMD/W 60/70 and
HMP140 series analogue transmitters
3 = calibrator for HMP230 transmitters

4 = calibrator for HMD/W20/30 and
HMP130 series transmitters

 (number up) or

 (number down)

ON/OFF

Содержание HMD20

Страница 1: ...USER S GUIDE Calibration of Series HMD W20 30 and HMP130 Transmitters with Vaisala HUMICAP Indicator HMI41 U220EN11...

Страница 2: ...echanical including photocopying nor may its contents be communicated to a third party without prior written permission of the copyright holder The contents are subject to change without prior notice...

Страница 3: ...20en 1 1 Operating Manual 1996 08 29 i Contents 1 INTRODUCTION 1 1 1 HMI41 with calibration option 1 1 2 Selecting the HMI41 calibrator function 1 2 CALIBRATION 3 2 1 Getting started 3 2 2 Calibration...

Страница 4: ...HMI41 Calibration Option Operating Manual 19165ZZ U220en 1 1 ii 1996 08 29 This page intentionally left blank...

Страница 5: ...libration options please refer to corresponding manuals Calibration is quick and easy to perform In calibration the HMI41 measures relative humidity with the reference probe and receives the RH inform...

Страница 6: ...librator for HMD W 60 70 and HMP140 series analogue transmitters 3 HMI41 as a calibrator for HMP230 series digital transmitters 4 HMI41 as a calibrator for the HMD W20 30 and HMP130 series analogue tr...

Страница 7: ...may vary from 10 minutes to a couple of hours 2 1 Getting started After having selected the desired function Chapter 1 2 turn the HMI41 off and connect the calibration cable to the EXT connector at t...

Страница 8: ...h the ON OFF button and wait until a text similar to the following appears on the display RH Numbers on the first line indicate the RH measured by the transmitter and numbers on the second line indica...

Страница 9: ...ference probe If you wish you can now calibrate the transmitter by pressing MODE again the measured values appear on the display Wait until the readings have stabilized and adjust the calibration pote...

Страница 10: ...unction press ON OFF until you can see some text on the display Then release the ON OFF button and press within 1 2 seconds both ENTER and MODE buttons until the text setup appears on the display Then...

Страница 11: ...ge RH Let the readings stabilize and then adjust the transmitter potentiometers until the readings are the same offset 65 RH gain 65 RH Alternatively you can adjust the transmitter when the display sh...

Страница 12: ......

Страница 13: ......

Страница 14: ...www vaisala com U220EN...

Отзывы: