HMI41 Calibration Option
19165ZZ-U220en-1.1
Appendix 1
1996-08-26
1
APPENDIX 1: QUICK REFERENCE TO CALIBRATION
It is recommended that this quick reference guide is used only as a checking
list for those who already know how to operate the HMI41 as a calibrator. For
those who take it into use for the first time, the Operating Manual gives useful
information that is not included in this quick reference guide.
1.
Selecting the calibrator function
For selecting the calibrator function, press ON/OFF until you can see some
text on the display. Then release the ON/OFF button and press within 1...2
seconds both ENTER and MODE buttons until the text "setup" appears on the
display. Then follow the instructions of the table below.
DISPLAY
WHAT TO DO
PRESS:
Wait for a few seconds.
set
°C
If the basic settings have been given,
press ENTER.
If they have not been given, refer to the
HMI41 Operating Manual.
press ENTER repeatedly:
set
select the HMI41 function:
1 = indicator
2 = calibrator for HMD/W 60/70 and
HMP140 series analogue transmitters
3 = calibrator for HMP230 transmitters
4 = calibrator for HMD/W20/30 and
HMP130 series transmitters
▲
(number up) or
▼
(number down)
ON/OFF