![Vaisala Drycap DMT152 Скачать руководство пользователя страница 62](http://html1.mh-extra.com/html/vaisala/drycap-dmt152/drycap-dmt152_user-manual_3228362062.webp)
8. Technical Data
8.1 Specifications
Table 38 Measurement Performance
Property
Description/Value
Sensor
Vaisala DRYCAP
â
180U
Thin-film capacitive polymer sensor
Recommended calibration interval
2 years
Dew Point Temperature
Measurement range
−80 … −20 °C (−112 … −4 °F) T
d
Accuracy
−80 … −40 °C (−112 … −40 °F)
±2 °C (3.6 °F) T
d
−40 … −20 °C (−40 … −4 °F)
±3 °C (5.4 °F) T
d
Non-calibrated range
−100 … +20 °C (−148 … +68 °F) T
d
Typical response time 63 % [90 %] at a gas temperature of +20 °C (+68 °F) and pressure of 1 bar:
−20 … −80 °C T
d
0.5 min [7.5 min]
−80 … −20 °C T
d
2 s [5 s]
Typical long-term stability
Better than 2 °C (3.6 °F) / year
Analog Output Scalings
Option 1
−80 … +20 °C (−112 … +68 °F) T
d
Option 2
−100 … 0 °C (−148 … +32 °F) T
d
Option 3
Free scaling
Concentration by Volume (ppm)
Measurement range (typical)
0 … 500 ppm
Accuracy at +20 °C (+68 °F), 1013 mbar
±(0.2 ppm + 20 % of reading)
1)
When the dew point is below 0 °C, the transmitter outputs frost point for T
d
.
DMT152 User Guide
M210865EN-J
60