Syntax
Description
Example (adjust the internal temperature measurement to 23 °C at the current conditions):
pass
1300
ct
23
OK
Example (clear the offset adjustment from internal temperature measurement):
pass
1300
ct
-reset
OK
Table 34 CP command
Syntax
Description
CP
[pressure]
<cr>
Perform an offset correction to the internal
pressure measurement.
pressure
= Correct pressure in hectopascals
(hPa).
CP
[-reset]
<cr>
Clear the offset adjustment from internal
pressure measurement.
Example (adjust the internal pressure measurement to 990 hPa at the current conditions):
pass
1300
cp
990
OK
Example (clear the offset adjustment from internal pressure measurement):
pass
1300
cp
-reset
OK
More information
‣
Changing measurement settings (page 35)
Chapter 4 – Vaisala Industrial Protocol
43