User's Guide __________________________________________________________________________
12 _________________________________________________________________________ U421EN-1.1
PROBE_SAVE<
CR
>
>PROBE_SAVE<cr >
New calibration parameters are now saved in the probe.
14.
Disconnect the gas flow and remove the probe from the chamber.
After two-point calibration it is recommended to ensure the correct operation of
the probe by carrying out a spot checking procedure using both low and high end
gases. The readings should be within the accuracy specifications of the
transmitter.