6.4 Connect a breathing circuit to the chamber.
6.5
Connect temperature probe to humidifier, chamber and
inspiratory tube.
6.6 Connect the heated wire cable to the circuit.
Notes
1.DO NOT fill the chamber above the maximum fill level line.
Liquid could enter the breathing circuit if the chamber is
overfilled.
2.When mounting a humidifier adjacent to patient, ensure that
the humidifier is always positioned lower than the patient.
3.Ensure that both temperature probe sensor are correctly and
securely fitted. Failure to do so may result in temperatures in
excess of 41℃ being delivered to the patient.