- 9 - 

5. Performance Data

    

Follow the next steps to conduct product test: 

5.1  Put temperature tester on heater plate. 

5.2  Set temperature level as 3,check if the temperature is 

correspondent to the indicator on device panel. 

5.3  Set temperature level as 6,check if the temperature is 

correspondent to the indicator on device panel. 

5.4  Set temperature level as 9,check if the temperature is 

correspondent to the indicator on device panel. 

Suggested Tester: Fluke 572 Thermometer (Accuracy  ±1.0℃) 

Note: Please make performance test every 12 months. 

    

6. Disposition 

Please dispose humidifier according to the local rules of the 
disposition of electronic devices. 

    

7. Cleaning 

The Heater base may be wiped with a damp cloth with using clean 
water.

    

    

8. Maintenance 

8.1    Check the power cord for damage, replace if necessary. 
8.2    Check the heater plate for scratching, dirt or moisture; 

replace if necessary. 

8.3    Conduct cleaning when necessary. 

Содержание VH-1500

Страница 1: ...VH 1500 RESPIRATORY HUMIDIFIER Manual Read instruction carefully before use Made In Taiwan VA 1 800 VH1500 110 EN_01...

Страница 2: ...er 4 Setting up the humidifier 4 Setting up the humidifier 4 Setting up the humidifier 6 6 6 6 8 8 8 8 5 Performance 5 Performance 5 Performance 5 Performance Data Data Data Data 9 9 9 9 6 Disposition...

Страница 3: ...repair the device by yourself if it is malfunction Please contact local distributor for the service Must be used by qualified medical staff or trained personnel To avoid the risk of electric shock th...

Страница 4: ...er trap 2 Clean the parts with low flow running water using a soft brush to remove all foreign material for example soil or organic matter 3 Sterilization by autoclave 132 4minutes Filter 1 Take coupl...

Страница 5: ...d Compliance EN IEC 60601 1 EN 60601 1 2 Classified as Class I Type BF IPX1 Definitions of symbols Definitions of symbols Definitions of symbols Definitions of symbols TYPE B TYPE B TYPE B TYPE BF F F...

Страница 6: ...t length 22mmID 120cm VADI chamber G 314004 Power Level Heater plate Temperature Flow rate range LPM Temperature at patient Y 3 52 5 40 26 30 5 60 5 40 30 34 8 71 5 40 33 36 Temperature accuracy 3 Ove...

Страница 7: ...is button to cut off the alarm signal alarm continues signal after 2 minutes The scheme of power level shows the temperature of heater plate Caution Heater level is Stand by heater plate is not heated...

Страница 8: ...heater base Pic 2 Caution Choose the appropriate chamber 4 3 Add distilled water into chamber Pic 3 Or stab distilled water with the needle of IV SET and connect the tube to humidification chamber Pi...

Страница 9: ...in chamber may result water condensation in breathing circuit Do not touch the heater plate while operation Do not use any blanket to cover the breathing circuit Operating environment Suggested enviro...

Страница 10: ...erved during normal operation If there is no flow turn off the humidifier Water traps are recommended for use to collect water condensation due to differences in temperature Absorbent paper is recomme...

Страница 11: ...check if the temperature is correspondent to the indicator on device panel Suggested Tester Fluke 572 Thermometer Accuracy 1 0 Note Please make performance test every 12 months 6 Disposition Please di...

Страница 12: ...sable adult humidification chamber Reusable adult humidification chamber Reusable adult humidification chamber Reusable adult humidification chamber G G G G 314005 314005 314005 314005 Reusable infant...

Отзывы: