Prior to autoclaving set the VHC
pro
to “auto-
claving”
.
Prior to autoclaving turn cap (1) and remove
(2).
1
2
Cleaning and decontaminating
Autoclaving
The VacuuHandControl VHC
pro
with adapters
and aspiration hose is designated for steam
sterilization at 121°C and 2 bar absolute (1 bar
overpressure). Time of exposure according to
DIN 58946 t
e
= 20 minutes.
Ensure not to exceed a temperature of 121°C.
Comply with instructions for use of the manu
-
facturer of the autoclave.
It is the user´s responsibility to ensure effective
autoclaving.
NOTICE
121°C
2 bar absolute
NOTICE
It is not possible to disassemble the VHC
pro
completely.
It is possible to disassemble and replace the
adapters, the screw cap and the aspiration
hose, see “Accessories and spare parts”.
If the housing opens lightly by mistake, e. g.,
if the handle is fallen down, press the gap to
-
gether by hand. Perform a function test.
NOTICE