56
20901101_EN_PC5xx-6xx_select_Serie_V1.3_260819
Troubleshooting
Error
`
Possible cause
Remedy
Personnel
Venting valve
does not op-
erate
`
`
No voltage applied.
`
`
VACUU·BUS plug‑in
connection or cables
defective or not con-
nected.
`
`
Venting valve dirty.
`
`
Venting valve in sen
-
sor defective.
`
Check VACUU·BUS
plug‑in connection
and cables to the
controller.
`
Clean venting valve.
`
If necessary, use
another external
venting valve.
Specialist
Vacuum
pump does
not start.
`
`
Overpressure in the
outlet (exhaust) line.
`
`
Condensation in the
vacuum pump.
`
Open up exhaust
gas line.
`
Ensure a clear pas-
sage.
`
Flushing: Oper
-
ate vacuum pump
briefly with open
suction nozzle.
Operator
`
`
Pumping unit
switched off.
`
`
Power plug not cor
-
rectly plugged in or
pulled out.
`
`
VACUU·BUS plug‑in
connection or cables
defective or not con-
nected.
`
Switch pumping
unit on using rocker
switch.
`
Check power supply
and cable.
`
Check VACUU·BUS
plug‑in connection
and cables to the
controller.
Operator
`
`
Motor overloaded.
`
`
Thermal protection
triggered.
`
Allow the motor to
cool down.
`
Clear error manu-
ally:
Unplug pumping
unit from the power
supply
Eliminate cause
of error
Switch pumping
unit back on
Specialist
No or very lit-
tle suction
power
`
`
Leak in the suction
line or in the appara-
tus.
`
`
Separator flask not
mounted properly.
`
Check suction line
and apparatus for
possible leaks.
`
Check separator
flask and its cor
-
rectly fixation.
`
Check apparatus for
leaks.
Operator