34
20901101_EN_PC5xx-6xx_select_Serie_V1.3_260819
Installation and connection
Installing the vacuum pump
>
Place the vacuum pump on a stable, nonvibrating, level, hori
-
zontal surface.
5 cm
5 cm
5 cm
>
When installing in lab furniture, maintain a minimum distance
of 5 cm (2 in.) to adjacent objects or surfaces.
>
Prevent heat accumulation and ensure sufficient air circula
-
tion, especially in closed housings.
Observe limitations of use
Limitation of use
(US)
Ambient temperature
10
–
40 °C
50–104°F
Altitude, max.
2000 m
über NHN
6562 ft
above sea level
Minimum distance to
adjacent parts
5 cm
2 in
Coolant circuit, max. pressure
3 bar
44 psi
Relative humidity
30–85 %, non condensing
Pollution degree
2
Protection type/Impact energy
IP 40 / 5 J
Prevent condensation or contamination from dust, liquids, or corrosive
gases.
>
Note the IP protection class. IP protection is only guaranteed
if the device is appropriately mounted and connected.
>
For connection also note the rating plate data and chapter
8.1.1 Technical data on page 78
.
IMPORTANT!
Example
Sketch
Minimum distances
in laboratory
furniture
Observe limitation of
use
IMPORTANT!