Cleaning and maintenance
20901118_EN-US_PC301x VARIO select Serie_V1.3_100720
67
Replace valves
-> Example
Valve replacement
1. Take the pump head pair
which you had set aside.
2. Unscrew the Torx screws.
Torx screwdriver
size Tx20.
3. Remove the clamping
claws from the valve termi-
nals.
4. Remove the valve termi-
nals with the disc springs.
NOTICE!
n
The number and wiring of
molded hoses
(a)
depends
on the position of the pump
head pair. Pump head
pairs must not be inter-
changed.
n
Valves can adhere to the
underside of a valve termi-
nal.
Plan view example:
Valve terminals, valves, O-rings,
molded hose of a pump head pair.