Cleaning and maintenance
58
20901118_EN-US_PC301x VARIO select Serie_V1.3_100720
Replace pressure relief valve + O-ring
-> Example
Maintenance on the
suction/pressure
distributor
(only
PC 3010 or
PC 3012)
1. Place the vacuum pump
on a clean, stable surface
as shown.
2. Only open the hose clips
above the pressure distrib-
utor; flat-head screwdriver
size 1.
3. Remove the molded hoses
one by one from the hose
nozzles.
4. Unscrew the screw fittings.
Phillips screwdriver size 2.