page 21 of 96
In normal use, the lifetime of the diaphragms and valves
is typically 15,000 operating hours. Bearings have a typi-
cal durability of 40000 h.
➨
Ensure that the pump cannot be operated acciden-
tally. Never operate the pump if covers or other
parts of the pump are disassembled.
➨
Switch off the pump. Disconnect the electrical
power cord and wait two minutes
before starting
maintenance to allow the capacitors to discharge.
➨
Note
: The pump may be contaminated with process
chemicals, which have been pumped during operation.
Ensure that the pump is completely decontaminated
before maintenance commences.
+
Take adequate precautions to protect people from the
effects of dangerous substances if contamination has
occurred. Use appropriate protective clothing, safety
goggles and protective gloves.
+
Wear parts have to be replaced regularly.
+
Never operate a defective or damaged pump.
+
Vent the pump before starting maintenance. Isolate the
pump and other components from the vacuum system.
Allow sufficient cooling of the pump. Drain condensate,
if applicable.
Ensure that
maintenance
is done only by suitably trained
and supervised technicians. Ensure that the maintenance
technician is familiar with the safety procedures, which re-
late to the products processed by the pumping system.
Only dismantle the pump as far as necessary.
NOTICE
Содержание ME 16C NT VARIO o.C.
Страница 30: ...page 30 of 96 ME 16C NT VARIO o C MD 12C NT VARIO o C MV 10C NT VARIO 9 1 2 8 16 16 5 3 6 4 7 7...
Страница 54: ...page 54 of 96 3 2 4 4 4x...
Страница 55: ...page 55 of 96 6 5 1 2 2 5 mm 2 5 mm ME 16C NT VARIO o C MD 12C NT VARIO o C MV 10C NT VARIO o C 7...
Страница 56: ...page 56 of 96 2 5 mm Emission condenser Peltronic 8 6A Emission condenser Peltronic 4 2x 7A...
Страница 57: ...page 57 of 96 Emission condenser Peltronic 4 4x 9 8A 10 4 2x...
Страница 58: ...page 58 of 96 12 11 4 4x Don t loosen this screw 12A 13...
Страница 59: ...page 59 of 96 5 2x 12A 1 2 3 1 4 1 4 3 2x 5 13 14...
Страница 60: ...page 60 of 96 17 5 2x 16 15...
Страница 62: ...page 62 of 96 18 2 5 mm Replacing the diaphragms 19 20 Clean 5 8x 20696867 20696839...
Страница 63: ...page 63 of 96 21 22 23 1 2...
Страница 64: ...page 64 of 96 24 3 25 Pay attention to number of washers 26...
Страница 67: ...page 67 of 96 33 34 35 Clean...
Страница 68: ...page 68 of 96 36 37 38 IN IN EX EX 4x 20696867 20696839...
Страница 69: ...page 69 of 96 39 40 41 IN EX T20 4x 2 2 ft lbf 3 Nm...
Страница 70: ...page 70 of 96 42 43 44 Maintain all pump heads in the same way...
Страница 74: ...page 74 of 96 2 size 2 2x 3 4A Version A valve 24 4 mm...
Страница 75: ...page 75 of 96 5A 20696839 24 4 mm 26 mm 4B 5B 20696839 37 mm 40 mm Version B valve 37 mm...
Страница 76: ...page 76 of 96 6 7 size 2 2x 7A ME 16C NT VARIO o C MD 12C NT VARIO o C MV 10C NT VARIO o C size 2 2x...
Страница 78: ...page 78 of 96 Assembling the pump housing 46 5 2x 8 9 ft lbf 12 Nm 47 48...
Страница 79: ...page 79 of 96 49 5 2x 8 9 ft lbf 12 Nm 50 51...
Страница 80: ...page 80 of 96 52 52A 53 52A 1 2 3 3 2x...
Страница 81: ...page 81 of 96 53 4 4x 3 7 ft lbf 5 Nm 54 4 2x 3 7 ft lbf 5 Nm 55 4 4x 3 7 ft lbf 5 Nm...
Страница 82: ...page 82 of 96 56 57 58...
Страница 84: ...page 84 of 96 59 60 4 4x 3 7 ft lbf 5 Nm 61...
Страница 87: ...page 87 of 96 3 4 20638821 20638821 5...
Страница 88: ...page 88 of 96 6 7 T20 4x size 1 4x 8...