background image

1.  The tubing must be reversed.  C+ tube must now go to C- and C- must go to C+.   
 

So for double acting actuators you do exactly that, just reverse the tubing, for single acting, 

 

C+ would now be plugged and C- would now be piped to the actuator.  

2.  Remove the front cover and the indicator in order to expose the cam nut and cam-notice 
 

the cam position.

3.  Loosen the cam nut and take the cam out and flip it over-literally 180 degrees.  
 

Place the cam back in the positioner making sure that it is on the opposite lobe side. 

 

(example: on a C1 cam it would go from the C1 A 90D side to the C1 B 90R side)  

 

DO NOT TIGHTEN THE CAM NUT

.

4. 

a.

 is for double acting applications, 

b.

 is for single acting applications

a.  For double acting applications apply supply air to the positioner.  This should move the valve 
 

open, the desired 4mA position.  Now simply tighten cam nut.  Now the positioner will be 

 

working 4mA Open to 20mA close.  

b.  Since the positioner was set up originally 4mA closed we must now apply air to the 
 

actuator to get the valve 

 

to move to open.  This is 

 

done by pushing up on the 

 

balance arm. The 

 

balance arm is labeled 

#2

 

 

in this picture.  As you can 

 

see the balance arm is 

 

connected to the spool 

 

valve 

(#6)

 through 

 

connection 

#7

.  By pushing 

 

up on the balance arm we 

 

are exhausting air out of the 

 

C- port.  Once the valve is

 

open continue to hold the 

 

balance arm up and tighten 

 

the cam nut.  Once the cam 

 

nut is tightened release the balance arm, the positioner is now set to operate from 

 

4mA open to 20mA closed.

The positioner should now be operating as 4mA is open and 20mA is closed.

Note:  This will not change the “fail safe” action on an actuator with springs (usually 

single acting application).  To change the fail safe action you should contact the actuator 

manufacturer.

 

Отзывы: