Vacuum/Atmospheres
Table 5-1. VALVE ( SOLENOID AND MANUAL) FAILURE INDICATION (Cont.)
VALVE
VACUUM (K)
(continued)
FAILURE INDICATION
INotel
Regeneration failure is a term
which means that after a full
regeneration cycle (paragraph
3-5.1) the purity of the system is
not being maintained. Moisture and
oxygen are not being removed from
the saturated purifier. Therefore,
the purifier cannot perform its
function of removing oxygen and
moisture from gas circulating
through it.
- The purifier will not be evacuated during the regen
eration process, and the waste products generated as
moisture during earlier portions of the cycle wi 11
not be removed.
- The purifier remains saturated with moisture.
FAILURE TO DEENERGIZE
- If VACUUM valve (K) sticks - remains open - when the
purifier is refilled with inert gas from the gl ave
box, the box pressure becomes negative.
If pressure control in the gl ave
box system is turned off, or the
box runs out of makeup gas, the box
could be damaged.
5-8
HE-493-7/80