1.

 

Remove Carriage Clamp 

The PT-55 is shipped with a 

Carriage Clamp

 

(1)

 

attached to the carriage assembly to 

prevent carriage movement during shipment. Use the supplied 1/8” Allen Wrench to remove 
the (4) BHCS which hold the clamp in place. Remove the clamp and discard. 

If the optional laptop security bracket was purchased, visit the V-TEK, Inc. website at 

www.vtekusa.com

 for installation instructions. 

 

 

2.

 

Connect Power Supply 

Connect the AC power cord into the 

AC Power Receptacle

 

(2a)

 

which is located on the 

back of the PT-55. Connect the power cord to a 120/230 VAC power source. 

Flip the 

Power Switch

 

(2a)

 

which is located above the Power Receptacle to the ON 

position.

 

 

3. 

Configure HMI Settings 

Connect the PT-55 laptop to the 

USB Port

 

(3)

 

on the back of the PT-55, and then power it 

up.  

Note:

 If V-TEK optional computer was not purchased, load software onto user-supplied 

computer using provided USB Drive. See online manual for details. 

In the HMI 

Main Screen

, press 

Supervisor

 and enter the password at the prompt. The 

default password is “password”. Enter desired settings for 

Auto Tension

Save Mode

 and 

test parameters.

 

 

4.

 

Prepare Tape Sample 

Run out a 6 to 12 inch strip of empty, sealed carrier tape for testing. Orient the tape sample 
so it is right-side up with the sprocket holes facing the Peel Force Tester. At the right end of 
the tape, carefully peel back about 1 inch of cover tape. Attach the 

Alligator Clip

 

(4)

 

from 

the load cell chain to the loose cover tape. Make sure the clip is straight and centered on the 
tape.

 

 

5. 
 

 
 

6. 

Feed Tape 

Adjust the carriage track width, loosening the 

Track Adjustment Knob

 

(5)

 

as needed. The 

goal is for the track to touch the sides of the tape but still allow it to move freely.  

Advance the tape until it reaches the red 

Idler Wheel

 

(6a)

.

 Feed the carrier tape under the 

tape guide, and then lift the idler wheel to engage the sprocket holes of the carrier tape with 
the sprocket wheel teeth. Lower the idler wheel

Slide the 

Tape Adjuster Guide

 

(6b) 

forward until the edge of the carrier tape is in the groove of the tape guide. Tighten the 

Track Adjustment Knob

 

(5)

.

 

 

7. 

Position Load Cell 

Pull the 

Load Cell Assembly

 

(7a) 

out to a position where the chain is centered on the 

carrier tape. Adjust the 

Load Cell Position Thumb Screw

 

(7b) 

on the left side of the 

machine as needed.  

 

 

8.

 

Run Test 

Press the 

Start Test

 button in the HMI’s 

Main Screen

. The PT-55 will begin moving the 

carriage and graphing the results. The test will stop automatically when it reaches the preset 
pull length. You may also press the 

Stop Test

 button when enough data has been gathered 

to complete the test. The 

Main Screen

 will then display a 

PASS

 

or 

FAIL

 

message. 

 

 

 