measures the value 

L

100 LTR, current k value ( not to amend before )

0

k

=40, Calculated as follows 

0

I = K

0

L

0

K=

Users to manually modify the k value, for example 

Clear "0" TTL1 value or remove the batteries, to clear TTL1, TTL2 "0", fitted with

batteries, button is shifted to TTL1 status, start the measurement of liquid flow,

the flowed liquid enters into the standard cylinder. Record the displayed flow

value and graduated cylinder flow value after the liquid flows out, such as

displayed flow 

l

=95.56 LTR, graduated cylinder 

Note

 

l

= flow rate of flow meter 

k

= current flow

not to amend before

k value 

0

L

user graduated cylinder measures the value 

0

95.56x40

100

K=

38.224

As it only keeps the 2 numerical after two decimal, 

k

38.22, to modify 

value

to 38.22 is okay 

IV. Maintenance 

The computerized electronic meters can work at least 9,000 hours (one year) as

the battery power. In most cases, the batteries are required to be replaced for one

year. Prior to the demolition of the battery on sealed meter will extend batteries life.

If the meter readings are dim or blanking, the battery should be replaced. PLS use

batteries for Replacement from your distributor or factory. 

When the batteries are disconnected or damaged, circuits in batches and the total

cumulative return to zero. Factory on-site check at the time still without power

at the instrument computer. 

I strongly recommend checking the batteries and electrodes cleansing routine

should be used as part of the process of maintenance. Battery electrode should

be inspected once a year. No need to remove the instrument from the pipeline

system, the batteries can be replaced . 