In order to avoid the wrong reading caused by electric shock or personal injury,
instrument display appears
“
-
”
Symbol, you should replace the battery immediately.
To avoid electrical shock or personal injury, turn off the battery cover before replacing
the battery, and check that the test leads have been disconnected from the measuring
circuit.
Follow the steps below to replace the battery
1) Turn off the instrument power supply.
2) Remove all test leads from the input socket.
3) Unscrew the captive screws of the battery cover and remove it.
4) Remove the battery cover.
5) Replace the new 6F22 9V battery
6) Install the battery cover as it is, tighten the screws.
Note
:
The polarity of the battery can not be installed.
5.2 Replace the probe
Warning
When changing the test leads, you must replace the same
or the same grade of the test leads. Table pen must be intact,
table pen level: 1000V 10A.
If the pen insulation is damaged, such as wire wire exposed, you must replace the test pen.
6. Attachment
1
)
Table Probe
Grade
:
1000V 10A
1pc
2
)
Manual
1pc
4
)
Battery
6F22 9V
1pc