4.
The M12 5 pins connector on the rear side named “PLC” allows a PLC analogue communication. The
control unit waits for a signal in order to allow UV light emission.
If this step is not realized, the UV LED
lamp will not switch on.
•
If you are using manually the control unit, plug the M12 connector named “Schunt”.
•
Otherwise, please refer to the electrical scheme below to install an external communication (PLC).
Pin number
Color
Type
Function
1
Brown
out
+ 24 VDC
2
White
Out
0V
–
GND
3
Blue
In
0 VDC
UV emission not allowed
24 VDC
UV emission allowed
4
Black
-
START TIMER +
5
Blue
-
START TIMER -
M12 connector scheme on the left and
‘’Schunt” picture on the right
5.
Connect the power cable into the IEC C14 connector and into the wall. Then, the power switch can be
turned on.
6.
The user can adjust the UV light intensity from 20% to 99% of maximal power thanks to the potentiometer
(digits appear 20 to 99).
7.
The UV LED lamp is now ready to be used. Two operating modes are available:
•
ON/OFF mode
•
TIMER mode
➔
You can switch between these two modes thanks to the switch button on the front side.